PHOTOMETRIC
Optical Measurement Range: -0.20 to 4.0 Absorbance units (A)
Photometric Accuracy: +/-1
Stability: Drift of no more than 0.005A in 8
Light Source: Remote halogen lamp source with lamp saver feature
Wavelengths: Standard: 405, 450, 492, 545, 600, and 630nm
6-filter: 405, 450, 492, 545, 600, and 630nm
Filter options from 405nm to 700nm
Filter type: IAD hardcoat interference, 10nm half bandpass
Vessel: Standard 96-well micro titer plates, or strip trays.
ELECTRONIC:
Power Requirements: 115V or 230V AC, 50-60Hz (auto-sensing)
PC Connection: USB Port
Microprocessor: eZ80
Speed: Reads absorbance of 96 wells in about 8 seconds.
Plate Transport: Stepper motor
Certifications: Pending
SOFTWARE
Calculation Modes: Single point calibration, point-to-point curve fit,
polynomial regression, linear and sigmoidal regression. (log & linear) cut off absorbance, multipoint % absorbance. User-programmable open system, selectable plate formatting, alphanumeric test naming, automatic interpretation options, duplicate well options, curve plotting and editing, flags and error messages.
OTHER
Enclosure:
Flame-retardant ABS plastic cover and base
Dimensions: Approx. 7.5 x 13.0 x 16.5 (19 x 20.3 x 42cm)
Weighs 10lbs. (4.5kg)
There are no reviews yet.