Key Features
o I measuring channel
o Automatic start with standard pipettes
o Increased throughput by additional preparation position
o RFID for security of use
o Resistive colour touch-screen
o Built-in thermal printer
SPECIFICATION
Principles: :Clotting assays; Light Scatter Principle
(640nm)
: Immunogenic Turbidimetric Assays (800nm
Test Menu: PT,APTT, FBG, Intrinsic and Extrinsic Factor
Assays, LA, PS,PC and D Dimer
Number of Channels: One Channel
Reaction Cuvette Holder: 10 Positions for incubator of single reaction on
cuvette
Reagent Cup Holder: 5 Positions (4 cups and 1 bottle with stirring
programmeable)
Display: Large graphical touch screen display
Printer: Built-in Graphics Thermal Printer
Units of Measurement: seconds, %, INR, mg/dl,gl/dl, ratio, etc
Coagulation Clotting Curve: Yes
Operating Temperature and Humidity: 17-32 degrees Celsius, max 80% RH, non-
condensing
QC: QC with L-Jgraph
Power Requirement: 100-250V47-63Hz, 150VA
Interface: USB port, Pen drive/US compatible
Power Consumption: 150VA or less
Dimension and Weight: 216 x 205 x 75mm, 0.9kg